TECHNICAL SPECIFICATIONS
Quality Standards and Classification
- Anti-electroshock degree:
Class I equipment
[Internal power supply]
TEMP/SpO₂/NIBP: BF
ECG/Resp: CF
Application Range
- Adult/Pediatric/Neonatal/Medicine/Surgery/Operating Room/ICU/CCU
Display
- 14.0-inch TFT LCD screen
Resolution: 1366×768
Environment
Working Environment Conditions
- Temperature: 0–40°C
- Humidity: ≤85%
Shipping and Storage Conditions
- Temperature: -20–60°C
- Humidity: ≤93%
Electrical Specifications
- Power supply voltage: AC 100–240V, 50/60Hz
- Maximum input power: 70VA
- Fuse: F2AL 250V
- Working method: Continuous operation
- Built-in lithium battery: DC 11.1V, ≥2000mAh; working time ≥2 hours
Dimensions and Weight
- Product size: 336×262×99 mm
- Product weight: 2.6 kg
ECG
- Gain: ×0.25, ×0.5, ×1, ×2, Auto
- Waveform speed: 12.5 mm/s, 25 mm/s, 50 mm/s
- Sensitivity: >200 μV (peak-to-peak)
- Bandwidth:
- Diagnostic mode: 0.05–130 Hz
- Monitoring mode: 0.5–40 Hz
- Operation mode: 1–20 Hz
- Common Mode Rejection Ratio (CMRR):
- Diagnostic mode: >90 dB
- Monitoring mode: >100 dB
- Operation mode: >100 dB
- Electrode polarization voltage: ±300 mV
- Input impedance: ≥5 MΩ
- DC bias voltage: ±300 mV
- Calibration signal: 1 mV (peak-to-peak), accuracy ±5%
- ST segment detection:
- Measurement range: -2.0 mV to +2.0 mV
- Accuracy: ±0.02 mV or ±10% (whichever is greater) within -0.8 mV to +0.8 mV
Trend Data Length
- Alarm limit list: 400 groups
- Waveform playback: 6000 seconds
- Trend table: 7 days
- Trend graph: Up to 7 days
- Alarm review: 60 groups
ST Segment
- Maximum upper limit: +2.0 mV
- Minimum lower limit: -2.0 mV
- Adjustment increment: 0.1 mV
NIBP (Non-Invasive Blood Pressure)
- Measuring range and precision:
- Adult:
- Systolic: 40–270 mmHg
- Diastolic: 10–215 mmHg
- Mean: 20–235 mmHg
- Static pressure range: 0–300 mmHg
- Static pressure precision: ±3 mmHg
- Dynamic pressure precision:
- Max. average error: ±5 mmHg
- Max. standard deviation: 8 mmHg
SpO₂
- Measuring range: 0–100%
- Resolution: 1%
- Accuracy:
- ±2% (70–100%)
- Undefined (0–69%)
Pulse Rate
- Range: 30–300 bpm
- Resolution: 1 bpm
- Accuracy: ±3 bpm
- Update time: 1 second
Heart Rate
- Adult: 15–300 bpm
- Neonatal/Child: 15–350 bpm
- Accuracy: ±1% or ±1 bpm (whichever is greater)
Respiration Rate
- Adult: 0–120 BrPM
- Child/Neonatal: 0–150 BrPM
- Accuracy: ±2 BrPM (1–120 BrPM); undefined outside this range
Temperature
- Range: 0–50°C
- Resolution: 0.1°C
- Accuracy: ±0.1°C (excluding sensor error)
Standard Accessories
- NIBP cuff & tube
- ECG cable & electrodes
- SpO₂ sensor
- Temperature probe
- Lithium-ion battery
- Power cable
- Operator’s manual
Optional Accessories
- End-tidal CO₂ (optional)
- Recorder (optional)
- Dual NIBP, Nellcor (optional)
Store & Review
- ECG: 30 minutes of single-lead ECG waveform storage
- Alarm: 1,800 alarm event records
- NIBP: 400 sets of NIBP measurement data
- Arrhythmia: 128 arrhythmia events (with 8-second ECG waveforms)
- Power-off storage: 72 hours of trend data + 1 ECG waveform (optional)
ECG Specifications
- Modes:
- Standard: 5-lead
- Optional: 3-lead or 12-lead configurations
- Selectable Leads: I, II, III, aVR, aVL, aVF, V1-V6 (optional)
- Gain Settings: AUTO, 0.25×, 0.5×, 1.0×, 2.0×, 4.0×
- Insulation Breakdown Voltage: 4,000V AC (50/60Hz)
- Sweep Speeds: 6.25 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s
- Heart Rate Ranges:
- Adult: 10–300 bpm
- Pediatric/Neonatal: 10–350 bpm
- HR Accuracy: ±1% or ±1 bpm (whichever is greater)
IBP (Invasive Blood Pressure, Optional)
- Channels: 2
- Measurement Range: –50 to +300 mmHg
- Units: mmHg or kPa
- Accuracy: ±2 mmHg or ±2% (whichever is greater)
Alarm System
- Priority Levels: Low, Medium, High
- Alert Types: Audible + visual indicators
- Configuration: Default presets or customizable settings
- Silence Function: All alarms can be temporarily muted
- Volume Range: 45–85 dB (measured at 1-meter distance)
Thermal Recorder (Optional)
- Technology: Thermal dot array printing
- Paper Width: 50 mm (1.97 inches)
- Paper Speeds: 12.5/25/50 mm/sec
- Maximum Traces: 3 simultaneous waveforms
EtCO₂ (Optional, Microstream Technology)
- Measurement Range:
- Units: %, mmHg, or kPa
- Respiratory Rate Range: 2–150 breaths per minute (bpm)
Key Features Summary:
- Comprehensive Data Storage: ECG, alarms, NIBP, and arrhythmia records
- Flexible ECG Configurations: Supports 3/5/12-lead setups
- High Safety Standards: 4,000V AC insulation protection
- Customizable Alarms: Adjustable thresholds and priorities
- Optional Modules: IBP, EtCO₂, and thermal printing for expanded functionality