Gunson G4125 Gastester Digital Portable Exhaust Carbon Monoxide CO Analyser

Features and Benefits:

Recalibration :

The easiest way to do this is-

  • Centralise the adjustment knob on the front of the unit by twisting it all the way left then all the way right. You should be able to get it roughly in the centre from feel.
  • Turn the unit upside down and take out the screw, the unit should now split in half. *make sure you put the top half in an upright position or the tool will not calibrate correctly*
  • Gently take off the face plate covering the PCB *DO NOT ADJUST ANY OF THE POTS APART FROM THE ONE IT THE PICTURE!
  • The pot you are about to adjust does the same job as the adjustment knob on the front of the unit just at a greater value.
  • Connect to tool up to a 12v battery and let it warm up for 15 mins in fresh air, remember the top half of the unit has to be in the upright position.
  • After 15mins adjust the pot in the photo with a screwdriver until you get to 2.0%. *let the tool sit for a 5mins then adjust to 2.0% again*
  • Put the unit back together, your tool is now calibrated and you should be able to adjust and use the tool as per the instructions.