Features:
1. The steam sterilizer uses the advanced technology of 3 times pre-vacuum and vacuum drying. It ensures the hot steam to penetrate in thin tube or porous material.
2. The software program provides auto diagnose, protection against over heating or over pressurizing, and thus ensures reliability of the sterilization results.
3. Build- in condenser water collecting system prevents discharge of waste steam and keeps the machine clean and safe.
4. The real time display shows all the data during the sterilization cycle.
5. LED display panel. Double door locking-system. Sterilizing Temperature/Pressure(Under standard atmospheric pressure)121°C / (100-130) Kpa, 134°C / (200-240) Kpa.
Specification:
Chamber inner dimensions: 8L=17cm×32cm(6.7*12.6'')/12L=20cm×36cm(7.9*14.1'')
Rated voltage: 110V, 60Hz
Rated power: 1800VA
Sterilizing Temperature: 249.8℉ /273.2℉ (121℃/134℃)
Sterilizing Pressure: 100-130 Kpa/200-240 Kpa
Water tank capacity: 3.5 L
Environment temperature: 41℉~104℉(+5°C~+40°C)
Machine size: 41*56*42cm/16.1*22*16.5''
Packaging size: 68*51.5*53cm/26.7*20.2*20.8''
Transportation and Storage:
Temperature: -40 ~ +55 °C
Humidity: < 85%
Atmospheric pressure: 500Hpa -1060 Hpa
Package:
Tray rack *1
Drainage joint *2
Tray *3
Water filter net *1
Print paper *1
Draining filter net *1
Tray holder *1
Door gasket *1
Power cord *1
Water outlet tube *1
Caution:
1. If the front of the machine is low and the rear is high, it will affect the drainage and final drying effect. Please pay attention to properly raise the front foot of the machine.
2. The power supply must be able to afford 1800VA and grounded correctly.
3. Be sure to use distilled water.
4. The instruments to be sterilized should be placed in the trays without touching each other, allowing room for steam to pass through. It is suggested to using the tray holder to put the tray with the instruments inside the chamber to avoid scalding.
5. Do not open the door till the pressure is less than 9 kPa.
6. Turn off the power and let it cool down before you replace the door gasket, which could avoid any burning injuries.