Insulation resistance test |
Rated voltage |
UT513B |
UT513C |
Measurement accuracy |
|
250V |
0.25MΩ~2.49GΩ |
0.25MΩ~4.99GΩ |
±(5%+5) |
|
|
2.50GΩ~24.9GΩ |
5.00GΩ~49.9GΩ |
±(10%+10) |
||
|
500V |
0.50MΩ~4.99GΩ |
0.50MΩ~4.99GΩ |
±(5%+5) |
|
|
5.00GΩ~49.9GΩ |
5.00GΩ~99.9GΩ |
±(10%+10) |
||
|
1000V |
1.00MΩ~9.99GΩ |
1.00MΩ~9.99GΩ |
±(5%+5) |
|
|
10.0GΩ~99.9GΩ |
10.0GΩ~199GΩ |
±(10%+10) |
||
|
2500V |
2.50MΩ~24.9GΩ |
2.50MΩ~24.9GΩ |
±(5%+5) |
|
|
25.0GΩ~249GΩ |
25.0GΩ~499GΩ |
±(15%+10) |
||
|
5000V |
5.0MΩ~49.9GΩ |
5.00MΩ~49.9GΩ |
±(5%+5) |
|
|
50.0GΩ~499GΩ |
50.0GΩ~999GΩ |
±(15%+20) |
||
|
Short-circuit current |
Around 3.5mA(10S) |
Around 5.0mA(10S) |
||
|
Output voltage |
Rated voltage |
Step-able voltage range |
Display accuracy: ± (10% reading +10V) Output accuracy: + (0% ~ 20%) |
|
|
250V |
250V~300V |
|||
|
500V |
250V~600V |
|||
|
1000V |
500V~1200V |
|||
|
2500V |
1250V~3000V |
|||
|
5000V |
2500V~5000V |
|||
|
Dc voltage measurement |
30~1000VDC |
30~1000VDC |
±(3%+5) |
|
|
Ac voltage measurement |
30~750VAC;50Hz/60Hz |
30~750VAC;50Hz/60Hz |
±(3%+5) |
|
|
Capacitance measurement |
NA |
0.01uF~15.0uF |
±(15%+3) |
|
|
Leakage current display |
0.01nA~3.5mA |
0.01nA~5.0mA |
±(10%+5) |
|
|
communication |
USB communication |
USB communication /Bluetooth APP |
||
|
Special function |
||||
|
Power supply |
rechargeable lithium battery 14.8V 5200mAh |
|||
|
The polarization index/absorption ratio is automatically displayed |
√ |
|||
|
Continuous measurement of insulation resistance |
√ |
|||
|
Timing measurement of insulation resistance/comparison measurement |
√ |
|||
|
Current display |
Insulation resistance measurement and current display synchronously |
|||
|
Live monitoring of the subject |
Monitor the voltage of the measured object. When the voltage is greater than 36V, no test is allowed to protect the instrument and the operator |
|||
|
Test timer |
Automatically record the test time, timing range: 0 seconds ~ 99 minutes 59 seconds |
|||
|
High pressure warning |
When the safety voltage is exceeded, the danger warning symbol blinks |
|||
|
Storage function |
Store test data, a total of 999 groups |
|||
|
Battery level display |
There is a battery power display, battery low reminding timely charging |
|||
|
Automatic shutdown function |
Automatically shut down after 15 minutes (without high voltage output) |
|||
|
Basic parameter |
||||
|
Instrument size |
238mm(L)x166mm(W)x90.5mm(D) |
|||
|
Body weight (with battery) |
1800g(including battery) |
|||
|
Test line |
Double ended red high voltage test line 1, green test line 1, black test line 1 |
|||
|
Standard packing |
Toolbox + cartoon box |
|||
|
Standard packing quantity |
1 PC |
|||
|
Standard fittings |
User manual *1, Quick guide *1, test lead *3, USB cable *1, lanyard *1, charger: 1, Toolbox + cartoon box |
|||
|
Working environment |
0℃ ~ 35℃; < 75%RH, non-condensing |
|||
|
Storage environment |
-20℃~60℃;<80%rh |
|||
|
Pressure resistance |
CAT Ⅳ 600V |
|||
|
Fit for safety |
EN 61010-1ENIEC61010-2-034 BS EN61010-1BS ENIEC61010-2-034 |
|||