TECHNICAL SPECIFICATIONS
Multi-parameter Patient Monitor
Quality Standards and Classification
· CE, ISO13485

· SFDA: Class IIb

· Anti-electroshock degree: Class II (internal power supply)

· TEMP/SpO2/NIBP: BF

· ECG/Resp: CF

Application Range
· Adult/Pediatric/Neonatal/Medicine

· Surgery/Operating Room/ICU/CCU

Display
· 12" real color TFT screen

· Resolution: 800*600

· One alarm indicator(yellow/red)

· One working indicator(green)

· One battery charge status indicator(green)

· Three modes in accordance with the alarm state

Environment
Operating Environment
· Temperature: 0 ~ 40 °C

· Humidity: ≤85%

· Altitude: -500-4600m

Transport and Storage Environment
· Temperature: -20~60 °C

· Humidity: ≤93%

· Altitude: -500-13100m

Power Requirements
· AC: 100 ~ 240V, 50Hz/60Hz

· DC: Built-in rechargeable battery

· Battery: 5200mAh 11.1V lithium battery

· 4h operating after full charge (one piece)

· 5min operating after low battery alarm

Dimension and Weight
Equipment
· 369mm178mm341mm (about)

· 4.5kg (about)

Package
· 444mm426mm352mm (about)

· 6.5kg (about)

Data Storage
· Trend diagram/table: 7x24h

· NIBP review: 400 events

· Wave review: 100min

· Alarm review: 100 alarm events

· Support drug concentration titration analysis

ST Segment
· ST Segment Range: -2.0mV ~ +2.0mV

· Accuracy: 0.02mV

Pulse Rate
· Range: 20 ~ 300bpm

· Resolution: 1bpm

· Precision: ±3bpm

ECG
· 5 Leads: RA, LA, LL, RL, V

· Lead mode: I, II, III, aVR, aVL, aVF, V

· Increase: ×250, ×500, ×1000, ×2000

· Sensitivity: > 200 uV (Peak-to-peak value)

· Input impedance: > 5 (megohm)

· Bandwidth: Surgery 1 ~ 20 Hz; Monitor 0.5 ~ 40 Hz; Diagnostic 0.05 ~ 130 Hz

· CMRR: 100dB

· Polarization Voltage: ± 300mV

· Baseline Recovering Time: After defibrillation 3 seconds

· Signal Range: 8 mV (Peak-to-peak value)

· Calibrating Signal: 1mV, precision ±5%

· Heart rate range: 15-380bpm

· Heart rate accuracy: ±V-1%

RESP
· Method: RA-LL impedance

· Resp Impedance range: 0.3 ~ 3

· Base Impedance range: 200 ~ 4000

· Bandwidth: 0.1 ~ 2.5Hz

· Resp rate: Adult 0 ~ 120BrPM; Neonatal/Pediatric 0 ~ 150BrPM

· Resolution: 1BrPM

· Precision: ±2 BrPM

· Asphyxia Alarm: 10 ~ 40 seconds

NIBP
· Method: Pulse wave oscillometry

· Work mode: Manual/ Auto/ STAT

· Measure interval of auto mode: 1,2,3,4,5,10,15,30,60,90,120,180,240,480 (minutes)

· Measuring Time of STAT Mode: 5 minutes

· PR range: 30 ~ 250bpm

· Measure & alarm range:

o Adult: SYS 40 ~ 270mmHg; DIA 10 ~ 215mmHg; MEAN 20 ~ 235mmHg

o Pediatric: SYS 40 ~ 200mmHg; DIA 10 ~ 150mmHg; MEAN 20 ~ 165mmHg

o Neonatal: SYS 40 ~ 135mmHg; DIA 10 ~ 100mmHg; MEAN 20 ~ 110mmHg

· Static pressure range: 0 ~ 300mmHg

· Precision: ±3mmHg

· Pressure precision: Max. average error: ±5mmHg; Max. standard deviation: ±8mmHg

· Overvoltage protection: Adult 300mmHg; Pediatric 240mmHg; Neonatal 150mmHg

SpO2
· Range: 0 ~ 100%

· Resolution: 1%

· Precision: 70% ~ 100%: ±2 DIGIT; 0% ~ 69%: no definition given

TEMP
· Channel: 2

· Measure & alarm range: 0 ~ 50 °C

· Resolution: 0.1 °C

· Precision (no sensor): ±0.1 °C

IBP (Optional)
· Label: ART, PA, CVP, RAP, LAP, ICP, P1, P2

· Measuring and alarm range: ART 0 ~ 300 mmHg; PA 6 ~ 120 mmHg; CVP/RAP/LAP/ICP -10 ~ 40 mmHg; P1/P2 -10 ~ 300 mmHg

· Press Sensor: Sensitivity 5 uV/mmHg; Impedance 300-3000; Resolution 1 mmHg

· Accuracy: ±2% or ±1mmHg, which great

· Actualization interval: about 1 Sec

CO2 (Optional)
· Side/Main stream

· Warm-up time: when the ambient temperature is 25 °C, the carbon dioxide curve(capnogram) can be displayed within 20/15 seconds, and all the specifications can be fulfilled within 2 minutes.

· Measurement range: 0~150mmHg, 0-19.7%,0-20kPa(at 760mmHg), atmospheric pressure provided by the host.

· Resolution: 0.1mmHg: 0~69mmHg; 0.25mmHg: 70~150mmHg

· Precision: 0~40mmHg: ±2mmHg; 41~70mmHg: ±5% (reading); 71~100mmHg: ±8% (reading); 101~150mmHg: ±10% (reading)

· Respiratory rate range: 0~150 BPM

· Respiratory rate accuracy: ±1 BPM

Standard Accessories
· NIBP cuff & tube

· ECG cable & electrodes

· SpO2 sensor

· TEMP probe

· Lithium-ion battery

· Power cable

· Operator's manual

Optional
· Etco2

· CO2 modular

· IBP modular

· Touch screen

· Trolley bracket

· Hanging bracket

· Monitor recorder