HiPette 2uL-5000uL Mechanical Pipette Fully Autoclavable Adjustable Volume Lab
Fully autoclavable mechanical pipette for complete sterilization
Ultra-light, ergonomic design for comfortable, fatigue-free pipetting
Smooth, low-force plunger for effortless operation
One-key volume lock to prevent accidental volume changes
Precise and accurate pipetting with reduced operating force
Smooth knob volume adjustment with clear 4-digit volume display
UV-resistant and chemically corrosion-resistant materials
Easy disassembly for routine cleaning and maintenance
Single-Channel Adjustable Volume Pipettes
| Volume Range (μL) |
Increment (μL) |
Test Volume (μL) |
Systematic Error | Random Error | ||
|---|---|---|---|---|---|---|
| μL | % | μL | % | |||
| 0.1-2.5 | 0.002 | 2.5 | ±0.04 | ±1.4 | ±0.02 | ±0.7 |
| 1.25 | ±0.03 | ±2.5 | ±0.02 | ±1.5 | ||
| 0.25 | ±0.03 | ±12.0 | ±0.02 | ±6.0 | ||
| 0.5-10 | 0.01 | 10 | ±0.10 | ±1.0 | ±0.04 | ±0.4 |
| 5 | ±0.08 | ±1.5 | ±0.04 | ±0.8 | ||
| 1 | ±0.03 | ±2.5 | ±0.02 | ±1.8 | ||
| 2-20 | 0.02 | 20 | ±0.20 | ±1.0 | ±0.06 | ±0.3 |
| 10 | ±0.12 | ±1.2 | ±0.06 | ±0.6 | ||
| 2 | ±0.10 | ±5.0 | ±0.03 | ±1.5 | ||
| 5-50 | 0.05 | 50 | ±0.45 | ±0.9 | ±0.15 | ±0.3 |
| 25 | ±0.35 | ±1.4 | ±0.20 | ±0.8 | ||
| 5 | ±0.15 | ±3.0 | ±0.08 | ±1.6 | ||
| 10-100 | 0.1 | 100 | ±0.80 | ±0.8 | ±0.20 | ±0.2 |
| 50 | ±0.50 | ±1.0 | ±0.15 | ±0.3 | ||
| 10 | ±0.30 | ±3.0 | ±0.10 | ±1.0 | ||
| 20-200 | 0.2 | 200 | ±1.20 | ±0.6 | ±0.40 | ±0.2 |
| 100 | ±1.00 | ±1.0 | ±0.30 | ±0.3 | ||
| 20 | ±0.50 | ±2.5 | ±0.14 | ±0.7 | ||
| 100-1000 | 1 | 1000 | ±6.00 | ±0.6 | ±2.00 | ±0.2 |
| 500 | ±5.00 | ±1.0 | ±1.00 | ±0.2 | ||
| 100 | ±3.00 | ±3.0 | ±0.60 | ±0.6 | ||
| 1000-10000 | 10 | 10000 | ±60.00 | ±0.6 | ±20.00 | ±0.2 |
| 5000 | ±40.00 | ±0.8 | ±10.00 | ±0.2 | ||
| 1000 | ±30.00 | ±3.0 | ±6.00 | ±0.6 | ||
User calibration should refer to the industrial standard ISO8655-2.