• HiPette 2uL-5000uL Mechanical Pipette Fully Autoclavable Adjustable Volume Lab 


  • Fully autoclavable mechanical pipette for complete sterilization

  • Ultra-light, ergonomic design for comfortable, fatigue-free pipetting

  • Smooth, low-force plunger for effortless operation

  • One-key volume lock to prevent accidental volume changes

  • Precise and accurate pipetting with reduced operating force

  • Smooth knob volume adjustment with clear 4-digit volume display

  • UV-resistant and chemically corrosion-resistant materials

  • Easy disassembly for routine cleaning and maintenance


  • Mo ta anh

    Specifications

    Single-Channel Adjustable Volume Pipettes

    Volume Range
    (μL)
    Increment
    (μL)
    Test Volume
    (μL)
    Systematic Error Random Error
    μL % μL %
    0.1-2.5 0.002 2.5 ±0.04 ±1.4 ±0.02 ±0.7
    1.25 ±0.03 ±2.5 ±0.02 ±1.5
    0.25 ±0.03 ±12.0 ±0.02 ±6.0
    0.5-10 0.01 10 ±0.10 ±1.0 ±0.04 ±0.4
    5 ±0.08 ±1.5 ±0.04 ±0.8
    1 ±0.03 ±2.5 ±0.02 ±1.8
    2-20 0.02 20 ±0.20 ±1.0 ±0.06 ±0.3
    10 ±0.12 ±1.2 ±0.06 ±0.6
    2 ±0.10 ±5.0 ±0.03 ±1.5
    5-50 0.05 50 ±0.45 ±0.9 ±0.15 ±0.3
    25 ±0.35 ±1.4 ±0.20 ±0.8
    5 ±0.15 ±3.0 ±0.08 ±1.6
    10-100 0.1 100 ±0.80 ±0.8 ±0.20 ±0.2
    50 ±0.50 ±1.0 ±0.15 ±0.3
    10 ±0.30 ±3.0 ±0.10 ±1.0
    20-200 0.2 200 ±1.20 ±0.6 ±0.40 ±0.2
    100 ±1.00 ±1.0 ±0.30 ±0.3
    20 ±0.50 ±2.5 ±0.14 ±0.7
    100-1000 1 1000 ±6.00 ±0.6 ±2.00 ±0.2
    500 ±5.00 ±1.0 ±1.00 ±0.2
    100 ±3.00 ±3.0 ±0.60 ±0.6
    1000-10000 10 10000 ±60.00 ±0.6 ±20.00 ±0.2
    5000 ±40.00 ±0.8 ±10.00 ±0.2
    1000 ±30.00 ±3.0 ±6.00 ±0.6

    User calibration should refer to the industrial standard ISO8655-2.