Chromatography, MS & Filtration
Refrigerated HPLC autosampler with a 0.1 to 100 µL injection range and a 1290 Infinity thermostat for sample cooling.
| QC PASS |
Tested & Working
Lightly Used, Excellent
|
At a glance
| Injection range | 0.1 – 100 µL in 0.1 µL increments. Up to 40 µL with reduced injection volume kit [hardware modification required]. Up to 1500 µL with multiple draw [hardware modification required]. |
|---|---|
| Precision | <0.25 % from 5 – 40 µL, <0.5 % from 2 – 5 µL, <0.7 % from 1 – 2 µL, <1.5 % from 0.5 – 1 µL |
| Injection accuracy | 1 % [10 µL, n=10] |
| Pressure range | Up to 600 bar [8700 psi] |
| Sample viscosity range | 0.2 – 5 cp |
| Sample capacity | 2 x well plates [MTP] + 10 x 2 ml vials, 108 x 2 ml vials in 2 x 54 vial plate plus 10 additional 2 ml vials, 30 x 6 ml vials in 2 x 15 vial plate plus 10 additional 2 ml vials, 54 Eppendorf tubes [0.5/1.5/2 ml] in 2 x 27 Eppendorf tube plate |
| Injection cycle time | Typically <21 s using default conditions and injection volume of 5 µL |
| Carry over | Typically <0.004 % |
| Communications | Controller-area network [CAN], RS-232C, APG Remote: ready, start, stop and shut-down signals, optional four external contact closures and BCD vial number out |
| Safety features | Leak detection, safe leak handling, leak output signal for shutdown of pumping system, and low voltages in major maintenance areas |
| GLP features | Early maintenance feedback [EMF] for continuous tracking of instrument usage with user-settable limits and feedback messages. Electronic records of maintenance and errors. |
| Autosampler power | 100-240 V~, 300 VA, 50/60 Hz |
| Autosampler weight | 15.5 kg [35 lbs] |
| Autosampler dimensions [H x W x D] | 200 x 345 x 440 mm [8 x 13.5 x 17 inches] |
| Autosampler ambient operating temperature | 4–55 °C [39–131 °F] |
| Humidity | < 95 % r.h. at 40 °C [104 °F], non-condensing |
| Operating altitude | Up to 2000 m [6562 ft] |
| Safety standards | IEC, CSA, UL |
| Thermostat temperature range | Settable from 4 °C to 40 °C in 1 ° increments |
| Thermostat temperature accuracy [ambient < 25 °C, humidity < 50 %] | 3 °C to 8 °C at a setpoint of 4 °C |
| Thermostat temperature accuracy [ambient < 30 °C, humidity < 60 %] | 3 °C to 9 °C at a setpoint of 4 °C |
| Thermostat power | 100-120/220-240 V~, 260 VA, 50/60 Hz |
| Thermostat weight | 20.7 kg [46 lbs] |
| Thermostat ambient operating temperature | 4 – 40 °C [39 – 104 °F] |
Specs from manufacturer literature for reference. Verify against your application.