Digital Insulation Resistance Tester ETCR3500B/3520B/C Megohmmeter 5000V High Voltage 10KV
Specification:
Material: ABS
Color: As shown in the picture
Model: ETCR3500 (5KV), ETCR3500B(10KV)
Product Rated Voltage: 250V, 500V, 1KV, 2.5KV, 5KV, 10KV
Output Voltage Accuracy: (5%~10%) ±10V
Insulation Resistance Range: 0.5MΩ~2TΩ (5KV), 0.5MΩ~5TΩ (10KV)
Short Circuit Current: >5mA
Functions: Insulation resistance test (IR), polarization index test (PI), absorption ratio test (DAR), filter function (4 time selections 10S, 20S, 30S, 40S), voltage test (V), capacitance test ( uF), current test (nA).
Power Supply: Rechargeable lithium battery 5200mAhx1 (built-in battery)
Adapter: AC110~240V
Polarization Index Test: Support
Absorption Ratio Test: Support
Ramp Test Mode: Take 10% of the preset voltage as a step amount to test until the preset voltage
Filter Test Mode: 10S, 20S, 30S, 40S four options
Voltage Test: Range: AC/DC: 0V~1000V; accuracy ±5%rdg±3V
Capacitance Test: Range: 10nF~25uF; accuracy ±10%rdg±10nF
Current Test: Range: 0.01nA~5mA; accuracy ±5%rdg±0.5nA
Voltage Monitor: Monitor the voltage of the object under test, monitor the discharge state after the test, prohibit the test when the voltage is greater than 36V, protect the instrument and the operator
Test Timer: Automatically record test time, timing range: 0s~9999s
Storage Function: Upload the stored test data to the computer through the USB communication line.
Battery Power Display: There is battery power display, reminding to charge in time when the battery voltage is low
Automatic Shutdown Function: Automatic shutdown after 15 minutes of startup
Protection Level: Close the lid IP65, open the lid IP40
Working Environment: -20℃~50℃; 80%RH
Storage Environment: -25℃~65℃; 80%RH
Insulation Resistance: 50MΩ (1000V) (between test circuit and shell)
Withstand Voltage: AC 3Kv 50Hz 1min (between test circuit and shell)
How to Use:
Get the product, open the upper cover, the user can select the measurement range, rotate the selection, connect the measurement line, you can measure, there is a power line control next to the product screen, connect the adapter to charge. With a USB interface, data can be imported into the computer. The operation is simple and practical, and it is more comprehensive than the insulation resistance tester on the market.