General Introduction
This steam sterilizer is a precision instrument specializing in the operation occasion in need of simple, fast and efficient sterilization by the doctor or his/her assistants, and the doctor shall take the responsibility for such work.
This steam sterilizer is equipped with the microprocessor for intelligent control and man-machine interface for convenient operation. With the distinctions in displaying parameter and condition actively, evaluating malfunction automatically, shutting off for self-protection automatically in case of excessive temperature and pressure, this type steam sterilizer is reliable product in sterilizing.
This product is N-level sterilizer for dental, ophthalmology, operating room, laboratory, small dental equipment, surgical instruments sterilization 121 ℃/134 ℃ for rapid sterilization of items.
Tech. Parameter
Chamber size of the Steam Sterilizer (Diameter X Depth) 23L…249mmX460mm
Rated Voltage:220V 50Hz / 110V 60Hz
Sterilizing (Pressure/temp.):0.9~1.1Bar/121℃
2.0~2.3Bar/134℃
(Both of above are Under Standard Atmospheric pressure)
Rated Power:750VA
Fuse-Melting pipe:T15A
Operating Temp:0-40ºC
Material:Stainless steel #304
Inside capacity:16/18/23 Liter
SAFETY :Heating-over protection ; safety valve
Installation
The clearance between the exterior side surface of the steam sterilizer and the ambient neighboring articles shall be 10cm as retained, and the top surface shall be 20cm retained.
The air-cooling window attached to the exterior surface of the steam sterilizer shall be non-clog by dust or articles, and the steam sterilizer shall be located in the proper air flowing environment by recommendation of the supplier.
The steam sterilizer shall be installed on the horizontal working base.
Package Listing:
1.Main Machine 1 pc
2.Tray 3 pcs
3.Plug 1 pc
4.Braket 1 pc
5.Manual holder 1 pc
6.Fuse 2 pcs
7.Operation manual 1 pc
8.Measuring cup 1 pc
9.Sealing ring 1 pc
10.O ring of tap 2 pcs
11.Draining pipe 1 pc
Maintenance and Service
1 Disinfect and clean the interior surface of the steam sterilizer room by Medicinal alcohol once a month.
2 Replace the fuse-melter.
(1) Shut off the power
(2) Revolve the fuse base by counter-clockwise by screwdriver to bring out the melting fuse to be replaced;
(3) Replace the new fuse and reset the fuse base on the original position, then use the screwdriver to fasten the base by clockwise.
(4)Check the correctness of the parameter of the new fuse replaced.
3 Clean the sealing ring periodically
The user shall clean the sealing ring periodically for avoiding the influence to the seal caused by the dust/dirt left due to the long period use. Use the smooth cloth with distilled water to wipe the surface of the sealing ring or sealing cap gently. The user shall discharge the sealing ring for further cleaning or replacement if the air-leaking problem can not be resolved after the above process (The user shall discharge the sealing ring periodically for cleaning and inspection as recommendation by the supplier). Put back the ring to the original door groove after cleaning. Most Important: the inset work must be done equably to the groove During the installation, the laid four equal points of the ring must be inserted to the groove firstly, later for the same work for the remaining segments of the ring. After finishing, press the sealing ring equally by the force of hand.
4 Shut off the power before service or components’ replacement, and the service or replacement work shall be done by the supplier or his designated technician.