Keep the drug test cup, samples, and controls at 59-78 degrees.
Remove the cup from the sealed pouch and use it immediately.
β
Have the donor sign and date the cup label.
The donor puts urine in the drug testing cup and caps it. Start the timer now.
β Operator checks cap tightness.
β Remove the peel-off label.
β Check the temperature strip label 4-5 minutes after specimen collection. The green color band will appear to indicate the temperature of the urine specimen. Adulterated specimens should be 90-100 degrees F (32-38 degrees C).
β Colored bands in test strip results reflect drug test results. Read the result in 5 minutes. Test results after 8 minutes are invalid.
β To confirm positive results, send the cup and urine sample intact to a toxicology lab for confirmation.
COLLECT: Remove the urine drug test cup from the sealed pouch as
soon as possible and use it. Have the donor provide a urine sample.
WAIT: Verify the temperature of the specimen after 2-4 minutes
by examining where the green band emerges.
On the relevant temperature line, a green band will emerge.
READ: When youβre ready to read the results, remove the label.
The results of the tests should be read within 5 minutes. After 8 minutes, they are invalid.
Positive
Only one colored band appears in the control region (C)
No apparent colored band appears in the test region (T)
Negative
Two colored bands appear on the membrane.
One band appears in the control regions (C), and another in the test region (T).
Invalid
Any result without a control band at the specified read time must be disregarded.
Retest the procedure. If the problem persists, discontinue using the kit and contact your local distributor.