The ultra low-temperature freezer is provided to departments of medicine, health, chemical industry, scientific research, etc. for low-temperature freezing, storage, test.



Parameter description
SP - Shutdown temperature. When the temperature is lower than or equal to the shutdown temperature SP, the compressor will stop running (r1 to r2).
r0 -- return difference temperature. When the temperature is higher than the startup temperature (SP+r0), the compressor runs
r1 - Minimum shutdown temperature
r2 - The maximum shutdown temperature
P1 -- Temperature calibration. Used to compensate for measuring temperature
P4 -- Probe shows temperature with or without decimal point
A0 - Overtemperature alarm return error. When the temperature is lower than (high temperature alarm value -A0) or higher than (low temperature alarm value +A0), cancel the high temperature alarm
A1 -- Overtemperature alarm
High temperature alarm value = (SP+R0+A1)
Low temperature alarm value = (SP-A1)
A2 - The overtemperature alarm is prohibited during power-on. The alarm of overtemperature is prohibited during the A2 time when the power is just on
A8 -- Overtemperature alarm delay. When the temperature exceeds the limit, the timer will start. When the time reaches A8, the alarm will start
H5 - Advanced parameter access password
c0 -- Minimum downtime
c2: Compressor start-stop state when the probe is faulty
P4 - Indicates whether to display decimals
Parameter setting
Setting the temperature (SP) is the only parameter that can be set without entering a password.
Press the SET key.SP will appear on the screen.
Press SET again. The SP temperature is displayed.
You can use ↑ and ↓ to adjust the SP temperature
Press SET to confirm setting temperature.
Press the SET key and DOWN key at the same time to exit the parameter setting, or wait for one minute, the display will automatically exit the setting interface.
out prior notice.