Features:

1. The basic accuracy for measurement and output is 0.05%, display 5 digits.

2. Measurement function: DC voltage, thermocouple.

3. Output function: DC voltage, thermocouple.

4. 8 thermocouple types.

5. Auto cool-end compensation when output thermocouple.

6. C or F display.

7. 9V battery (not included), battery life is about 25 hours

8. Operation temperature: 0-50C; Humidity<=80%RH

9. Storage temperature: -10C to 55C; Humidity<=90%RH

9. Size: about 200x100x40mm

10. Weight: 550g

11. Accessories: test leads, PDF manual


Specifications:

 

Output

Range

Output range

Resolution

Accuracy

 

Voltage

100mV

-10.00 ~110.00mV

0.01mV

±0.05% Setting value +30uV

The max. output current ±5mA

1000mV

-100.00~1100.0mV

0.1mV

±0.05% Setting value ±0.3mV

Thermocouple

R

-40~1760

1

±0.05% Setting value ±3 (≤100) 

Adopt the ITS-90 temperature 

S

-20~1760

1

±0.05% Setting value ±2 (100)

standard (Note 1)

B

400~1800

1

±0.05% Setting value ±3 (400~600) 

  

±0.05% Setting value ±2 (600)

 

E

-200.0~1000.0

0.1

±0.05% Setting value 

  

±2 (≤-100)

 

K

-200.0~1370

0.1

 

 

J

-200.0~1200.0

0.1

±0.05% Setting value

 

T

-200.0~400.0

0.1

 

 

N

-200.0~1300.0

0.1

±1 (-100)

 

 

Measurement

Range

Output range

Resolution

Accuracy

Voltage

100mV

-10.00 ~110.00mV

10uV

±0.05% Measuring value ±30uV

Input Resistance 1MΩ

Thermocouple

R

-40~1760

1

±0.05% Measuring value ±3 (≤100) 

xInput tesistance1MΩ x 

S

-20~1760

1

±0.05% Measuring value ±2 (100)

Adopt the ITS-90

B

400~1800

1

±0.05% Measuring value ±3(400~600)

 

±0.05% Setting value ±2 (600)

temperature standard  

E

-200.0~1000.0

0.1

±0.05% Measuring value

(Note 3)

±2 (≤-100)

  

K

-200.0~1370

0.1

 

 

J

-200.0~1200.0

0.1

±0.05% Measuring value

 

T

-200.0~400.0

0.1

 

 

N

-200.0~1300.0

0.1

±1 (-100)

 

Note:

1. Not including the accuracy of inner temperature compensatory transducer.

The range of inner temperature compensatory transducer:-10-50°C, compensatory error ≤0.5°C.

2. Temperature coefficient±0.005%range/°C(0°C-18°C,28°C-50°