Precision AC Signal Recovery, Since 1985 - Still a Workhorse in Labs Today
Measure signals as small as nanovolts, buried in noise a thousand times larger. The SR510 uses a precision analog sine-wave multiplier for demodulation - clean, harmonic-free signal recovery without the spurious responses that plague square-wave designs.
| Parameter | Specification |
|---|---|
| Reference frequency range | 0.5 Hz – 100 kHz |
| Voltage input impedance | 100 MΩ, 25 pF (AC coupled) |
| Current input | 1 kΩ virtual ground, 10⁶ V/A gain |
| Full-scale sensitivity | 100 nV – 500 mV (voltage); 100 fA – 500 nA (current) |
| Input voltage noise | 7 nV/√Hz at 1 kHz |
| Input current noise | 0.13 pA/√Hz at 1 kHz |
| Dynamic reserve | 20 dB (Low) / 40 dB (Norm) / 60 dB (High); +20 dB with bandpass filter; up to 100 dB with line notch filters |
| Common-mode rejection | 100 dB, DC to 1 kHz (degrades 6 dB/octave above 1 kHz) |
| Time constants | 1 ms – 100 s (pre-filter, 6 dB/oct); 0, 0.1, or 1 s (post-filter, 6 dB/oct) |
| Phase resolution | 0.025° steps; phase noise 0.01° rms (1 kHz, 100 ms, 12 dB TC) |
| Signal filters | Line notch (−50 dB), 2× line notch (−50 dB), auto-tracking bandpass (Q=5) |
| Analog I/O | 4 A/D inputs, 2 D/A outputs, 13-bit resolution, ±10.24 V |
| Interfaces | RS-232 (standard), GPIB/IEEE-488 (optional) |
| Output | ±10 V BNC, <1 Ω output impedance |
| Power | 100/120/220/240 VAC, 50/60 Hz, 35 W max |
| Dimensions / weight | 17" × 17" × 3.5" (rack-mountable), 12 lbs |
Hall probe and strain gauge measurement, micro-ohm measurements, C-V testing in semiconductor research, low-level light detection, electron spin and NMR studies, and any experiment requiring recovery of small AC signals from a noisy background.
Your source for cost competitive test and industrial equipment on eBay