VICTOR 7200B Leakage protector Tester tests line AC voltage, leakage current, leakage action current, leakage current and leakage action time, test current from 5~1000mA (5,10,10,20,30,40,50,50,60,100,150,200,250,300,500,600,1000 mA), test leakage current action time range of 0~999ms, 854 × 480 full color touch LCD, with battery power display and set fault voltage reference Display, back brightness adjustment, internal wiring guidance diagram, convenient and intuitive, easy operation. Instrument with high precision, small volume, light weight, delicate appearance, easy to carry, is the ideal tool for field or indoor detection of various leakage protector.
Model comparison
Model Number | Leakage current gear | Leakage action time measuring range | AC voltage gear | Charging function | Features |
VICTOR 7200 | 15~500mA | 0~999ms | None | None | Portable type |
VICTOR 7200B | 5~1000mA | 0~999ms | 0~600V | Yes | Color touch screen, test current and doubling rate selection, phase angle selection |
Technical specifications
Function | Test the line AC voltage, leakage operation current, leakage inactive current and leakage operation time of various leakage protectors |
Power supply | 7.4V rechargeable lithium battery (full of approximately 8.4V) |
Test line voltage | Below AC 600V, under relay protection switch test; below AC 250V |
Rated voltage | AC 220V |
Leakage current gear | (5,10,15,20,30,40,50,60,100,150,200,250,300,500,600,1000)mA |
AC voltage gear | 0~600V |
Test the leakage action time range | 0~999ms |
Leakage action time display accuracy | 5%rgt±3%dgt |
Gear current accuracy | ±10%rgt±5dgt |
Gear voltage accuracy | ±2rgt%±3dgt |
Failure information indication | LCD shows "999" timeout indication, possibly tested rated action current gear too small or leakage protector fault |
Test line length | 1.8m |
Test mode | Manual or automatic |
Display mode | LCD:7-inch TFT color touch screen |
Instrument size | 215mm × 178mm × 83mm |
Battery voltage | When the battery voltage drops to 7.0V, indicate that the battery voltage is low, please charge in time. The data measured at this time is equally accurate. |
Working current | 365mA Max |
Working temperature and humidity | -10 ℃ ~40 ℃; below 80%rh |
Storage temperature and humidity | -10 ℃ ~60 ℃; below 70%rh |
Insulation and pressure resistance | 100MΩ or greater, between AC2kV/rms(housing and screws) |
Suitable for safety regulations | IEC1010-1, IEC1010-2-032, Pollution Level 2, CAT Ⅲ (600V) IEC61326 (EMC Standard) |