MASS-STREAM D-6320 MFM - Thermal Mass Flow Meter 65Lt/min CH4, IP65 protected

 

THE MASS FLOW METER HAS NEVER BEEN USED – BOXED WITH CALIBRATION CERTIFICATE

 

DETAILS

MFM D – 6230

SN 5105-6979D

FLOW 65Lt/min

GAS NATURAL (CH4)

WORKING P 0.050 BAR GAUGE

P MAX 10 BAR

POWER 24 VDC

OUT 0-5 VOLT

SEALS VITON

 

COMES WITH

INLET FLOW STRAIGHTENER PIPE – 240mm LONG / 10mm DIAMETER

OUTLET FLOW STRAIGHTENER PIPE – 240mm LONG / 10mm DIAMETER

ELECTRICAL CONNECTOR 5-PIN WITH A SHORT LENGTH OF WIRE

 

Proven direct inline CTA sensor (no bypass)

Compact, rugged design (IP65, dust and waterproof)

Very low pressure drop

Less sensitivity to humidity or dirt

 

Compact IP65 Mass Flow Meters for low flow rates of gases

Bronkhorst model D-6320 Mass Flow Meters (MFMs) are suited for precise measurement of flow at operating pressures between vacuum and 10 bar (Aluminium) or 20 bar (Stainless Steel). The MFM consists of a proven inline thermal (CTA) mass flow sensor and a microprocessor based pc-board with signal and fieldbus conversion and a PID controller for optional mass flow control by means of a separately mounted control valve. The instrument is IP65 compliant and can optionally be equipped with a modern, multi-functional and multi-colour display, with operator buttons on the instrument.

The digital MASS-STREAM series is characterized by a high degree of signal integrity and, as an option, up to 8 calibration curves of different gases and process conditions can be memorized in the instrument. In addition to the standard RS232 output the instruments also offer analog I/O. As an option, an on-board interface can be mounted to provide CANopen®, DeviceNet™, PROFIBUS DP, PROFINET, Modbus RTU or FLOW-BUS (other on request) protocols.

 

Measurement / control system

Flow range (intermediate ranges available)

min. 0,05…1 ln/min
max. 0,35…7 ln/min
(based on N2)

Accuracy (incl. linearity) (based on actual calibration)

± 1,0 % RD plus ± 0.5% FS (at calibration conditions)

Repeatability

< 0,2 % FS

Turndown ratio

1:20

Response time (sensor)

approx. 0,3 sec.

Operating temperature

0 … 50 °C

Temperature sensitivity

± 0,2% Rd/°C (Air)

Pressure sensitivity

± 0,3 %

Leak integrity, outboard

tested < 2 x 10-8 mbar l/s He

Attitude sensitivity

at 90° deviation from horizontal max. error 0,2 % at 1 bar typical N2

Warm-up time

30 min. for optimum accuracy,
within 30 seconds for accuracy ±4% FS

 

Mechanical parts

Sensor

Stainless steel SS 316 (AISI 316L)

Pressure rating (PN)

10 bar g for instrument body in aluminium,
20 bar g for instrument body in stainless steel SS 316

Process connections

G1/4" / compression type or face seal (VCR/VCO) couplings

Seals

standard: Viton®;
options: EPDM, Kalrez® (FFKM), FDA and USP Class VI approved compounds

Ingress protection

IP65

Instrument body

Aluminium AL 50ST/51ST (anodised) or stainless steel SS 316

Sieves and rings

Teflon or stainless steel SS 316

 

Electrical properties

Power supply

+15 … 24 Vdc ±10%

Max. power consumption

Supply
15 V
24 V

Basic model
115 mA
85 mA

Add. for fieldbus
80 mA
50 mA

Add. for display
30 mA
20 mA

Analog output

0…5 (10) Vdc or 0 (4)…20 mA sourcing output)

Digital communication

standard: RS232

 

Electrical connection

Analog/RS232

8 DIN (male);

PROFIBUS DP

bus: 5-pin M12 (female);
power: 8 DIN (male)

CANopen® / DeviceNet™

5-pin M12 (male)

Modbus RTU / FLOW-BUS

5-pin M12 (male)

Modbus TCP / EtherNet/IP / POWERLINK

bus: 2 x 5-pin M12 (female) (in/out);
power: 8 DIN (male);

EtherCAT®/ PROFINET

bus: 2 x 5-pin M12 (female) (in/out);
power: 8 DIN (male)

 

FOR MORE INFORMATION SEE

http://www.hoskinscientifique.com/uploadpdf/Instrumentation/M+W%20Instruments/hoskin_mass-stream_4b17d1bd8ad65.pdf

Please ask for the cost for posting outside the UK.