1. Pressure Calibration (Gauge Mode Example)
* Connect the device under test (DUT) to the DUT output port and the reference standard gauge to the standard gauge output port
* Rotate the fine adjustment pump counterclockwise to its starting position. Set the micro-adjustment knob to the mid-position
* Ensure the pressure release valve is fully closed
* Slowly rotate the fine adjustment handwheel clockwise to increase pressure until the desired calibration point is approached
* Use the micro-adjustment knob for precise fine-tuning to the exact pressure point
* To release pressure after calibration, open the pressure release valve
2. Vacuum Calibration
* Connect the device under test (DUT) to the DUT output port and the reference standard gauge to the standard gauge output port
* Rotate the fine adjustment pump clockwise to its starting position. Set the micro-adjustment knob to the mid-position
* Ensure the pressure release valve is fully closed
* Slowly rotate the fine adjustment handwheel counterclockwise to generate vacuum until the desired calibration point is approached
* Use the micro-adjustment knob for precise fine-tuning to the exact vacuum point
* To release vacuum after calibration, open the pressure release valve