Features
The
product uses saturated steam to rapidly and efficiently sterilize
articles. It can be used in hospitals, public health centers, health
stations, clinics in factories and mines, scientific research
institutes, etc, to sterilize medical equipment, surgical dressings,
glass utensils, solutions, substrata. It can also be used as cooker in
plateau areas and be used to make high quality drinking water in
industrial and mining enterprises.
1. Fully stainless steel structure.
2. Hand wheel type of quick-open door structure.
3. Indicator light indicates working state.
4. Double scale indication pressure gauge.
5. Time scope:0~60min.
6. Automatically shut off with beep reminding after sterilization.
7. Electric heated.
8. Easy to operate,safe and reliable.
9. With pressure steam adjuster, when pressure reached the setting value, it’ll keep and control.
10. With timer, when the time reached the setting, auto turn off power and alarm to tell sterilize end.
11.
With safety valve, release valve and gauge, when the steam pressure go
over 0.165Mpa, the release valve should exhaust the pressure to ensure
the safety, there is reading on the gauge face for both Mpa and
temperature for check the data.
12. With electric heating elements of immerge type, auto pressure control.
13. The container and drum adopted stainless steel 0Gr18Ni9Ti, anti-acid, anti-alkali and anti-rust.
Specifications
Voltage Rating: 2KW, AC 220V 50Hz (AC110V 60HZ)
Sterilization Temperature: 126℃
Temperature adjust range: 105℃ ~ 126℃
Timer range: 0~60min
Heat average: ≤±1℃
Sterilization Pressure: 0.14~0.16Mpa
Max.Safety pressure: 0.165Mpa
Sterilization Volume: 24L/18L
Dimension:410×410×630 mm( 24L)/410*410*520 mm(18L)
G. W / N. W:20/19 KG(24L) 19/18 KG(18L)
Characteristics
◆Fully stainless steel structure
◆Indicator light indicates working state
◆Double scale indication pressure gauge
◆Time scope : 0~60min
◆Automatically shut off with beep reminding after sterilization
◆Electric heated
◆Easy to operate,safe and reliable
Safety and maintenance
1.
This sterilizer is steam pressure vessel so do not impact in using and
forbid overpressure. If the pressure pointer of pressure gauge reaches
the 0.165MPa and the safety valve does not exhaust, it should stop using
at once and check up the safety valve.
2. The pointer should be at
zero position before usage the pressure gauge, otherwise the gauge has
been failure and it should repair or replace the gauge. The pressure
gauge should be regulated 1~2 times per year, ensuring sensitiveness and
correctness.
3. Drain the water inside the container and condensing
water below the sieve plate and had better wipe out water and deposit on
the electro-thermal tube in order to benefit the sterilization quality
and prolong the working life.
4. Gasket is easy to wear. If it does not wok well, see it whether become aging and hard, then replace it in time.