Features

Universal AC input / Full range
3 pole AC inlet IEC320-C14
Medical safety approved (2 x MOPP between primary to secondary)
Low leakage current <100uA
No load power consumption < 0.15W
High efficiency up to 94.5%
Class II power (with earth pin)
Protections: Short circuit / Overload / Over voltage / Over temperature
Fully enclosed plastic case
LED indicator for power on
Energy efficiency level VI
100% full load burn-in test
3 years warranty

General Specification 

AC input voltage range 80~264VAC ; 113~370VDC
AC inrush current (max.) Cold start, 120A at 230VAC
Overload protection 105%~135% rated output voltage, hiccup mode, auto-recovery
Withstand voltage I/P-O/P: 4kVAC, I/P-FG: 2kVAC, O/P-FG:0.5KVAC
Working temperature -30~+70°C (refer to output derating curve )
Safety standards ANSI/AAMI ES60601-1 , TUV EN60601-1 approved
EMC standards EN55011(CISPR11) class B, EN61000-3-2,3, FCC part 15 class B, EN61000-4-2,3,4,5,6,8,11,EN55024, EN61204-3,EN60601-1-2
Length of output cable 100cm of UL2464, 16AWG
Dimension (LxWxH) 220x 85x 46 mm
Standard DC plug R7B: Power DIN 4P with lock type