Optical Dissolved Oxygen Meter Fluorescent Dissolved Oxygen Gauge Range 0~20mg/L

 

Features:
Dissolved Oxygen meter with multiple parameters display: D.O.,PH and
temperature.
Signal outputRS-485
Support manual salinity compensation, air pressure compensation and
manual/auto calibration.
Application:
-Excellent to use in fresh water aqua farm.
-Excellent to use in sea water aqua farm.

TECHNICAL PARAMETERS

D O  Range

0~20mg/L or 0~200%

D.O. Accuracy

1%

Dissolved Oxygen Sensor Drift

< 1%per year

Sensor Material

POM and Stainless Steel

Temp. Range

0~40C

Temp. Accuracy

±0.2C

Measuring Distance

3m~5m

Sensor Cleaning Cycle

According to water condition and measuring frequency

Working Temperature

-20C~50C

Material Supplied

3v~6v

Power

<0 5w

Signal Output

RS-485

 

DIMENSIONS

        

OPERATIONAL INSTRUCTIONS

(1). KEY PAD OPERATION

+Increase the setting value and move up;     

 -Decrease the setting value and move down

<   Left shift               >   Right shift

menuOK/POWER ON/POWER OFF

(2) . OPERATION

POWER ON

Step1In the off state, connect the sensor first

Step2Press the menubutton and the screen

will light up to display the home page.

Step3Press the "+" button to start the test   and display the detected value.

Step4Press the "-" key to switch to the percentage of saturated dissolved oxygen.

POWER OFF

The first type      After the machine is turned on, the instrument will be turned off after 5 minutes by default. After the time is up to 4 minutes, the instrument will give an alarm. When the time is up to 5 minutes, the instrument will automatically shut down.

The second type   Manually adjustable automatic shutdown time, time mode is 5 minutes, 10 minutes, 30 minutes, 1 hour, 2 hours, never stop, immediately shut down. The instrument will issue a warning before the last minute of the timer, and will automatically shut down after 1 minute. (Except for the mode of never shut down and immediate shutdown)

The  third  type    Press  the  menu”  button  for  a  long  time  and  turn  off the  instrument immediately.

CALIBRATION

The first typeAutomatic Mode

Step1 : Entering the automatic calibration interface.

Step2Put the sensor in the air, press "Start" directly, and the instrument will automatically calibrate.

Step3After successful calibration, the word "Calibration Completed" is displayed.

The second typeManual Mode

Step1 : Entering the manual calibration interface.

Step2Put the sensor in the air, set K and B values, and press the "menu" button.

Step3   After successful calibration, display "Setup Completed". If the current calibration