Standard Resistor Calibration of Grounding Resistance Tester 0.001Ω~100KΩ




Product Description

1. The military grade wire-wound resistance is used for manufacturing, with stable performance and strong structure, especially convenient for calibration and testing.

2. Fully meet the national weather service order "lightning protection device detection qualification management approach" specified in the range of 10-3 ~ 105 Ω.

Technical Specification

Function

Used for calibration of grounding resistance tester, equipotential tester and other instrument

Range and Accuracy

0.001Ω;      2%

0.01Ω;       0.5%

0.1Ω;         0.2%

1.0Ω;         0.2%

10Ω;          0.05%

100Ω;        0.05%

1KΩ;          0.05%

10KΩ;        0.05%

100KΩ;      0.05%

Connection Wire Method

4-wire method

Power

1/2W

Temperature Coefficient

±25ppm

Dimension

173mmX125mmX70mm

Weight

360g

Test Line

Red 0.5m & black 0.5m, each with 2pcs

Working Temperature and Humidity

-10~40;80%rh below

Storage Temperature and Humidity

-20~60;70%rh below

Insulation Resistance

10MΩ above(between the circuit and shell 500V)

Withstand Voltage

AC 3700V/rms(between the circuit and shell)

Suitable Safety Standard

IEC61010-1,CAT  600V,Polluted Gradation