HP353 RF Power Meter Programmable attenuation value Saving function 0.5G~15GHz -10~+16dBm
Parameters:
1. Overview:
HP353 is a RF power meter, the meter through the button to complete the operation, the advantages: simple structure, easy to use, stable performance, reliable, high sensitivity and fast measurement speed.
2. Application:
● RF power measurement;
● Instrument calibration;
3. Technical parameter table:
No. Item Parameter Unit Remarks
1 Measuring power range -10 ~ +16 dBm (external RF attenuator can be extended to 100 DBM)
2 Measurement power resolution 0.1dBm
3 Measurement frequency range 0.5G ~ 15 GHz
4 Operating voltage 7 ~ 12 VDC
5 Working current <50 mA
6 Operating temperature -40℃ ~ 65℃ Celsius
7 Storage temperature -40℃ ~ 65℃ Celsius
8 Dimensions 55*45*10 mm long X wide X high
9 Weight 50g
4. Interface display and operation instructions:
(1) Unit description: dBm;
(2) Key description:
FUN: Function key; MODE: Set the interface. SEL: Self-adding or self-subtracting key;
4.1. Display measurement data
Interface display font:
First line "RF-POW-65.5 dBm"
Second line "110uV 0nW"
Note: Display signal dBm value, input voltage value, input power.
4.2. Set the RF power attenuation value
Interface display font:
First line "RF Attenuator"
Second line "SUB 00.0dB"
Key operation:
FUN: Function key; (Select ADD or subtract direction) SUB is displayed as minus and add is displayed as plus. (Long press the FUN key to save the set dB value)
SEL: Self-adding or self-subtracting key;
4.3, set the value of the measurement frequency (function: to correct the RF power of different frequencies)
Interface display font:
First line, "Measuring freq."
Second line "SUB 00.0GHz"
Key operation:
FUN: Function key; (Select ADD or subtract direction) SUB is displayed as minus and add is displayed as plus. (Long press the FUN key to save the set frequency value)
SEL: Self-adding or self-subtracting key;
Note:
Connect different RF power attenuators, set the corresponding attenuation coefficient to read the correct power parameters.
Five, wiring
P1 connection definition:
Serial number symbol meaning
1 Positive power terminal of the VDD module (7V to 12V)
2 GND Negative power terminal of the module
P2 connection definition:
Serial number symbol meaning
1 RF signal input (RP SMA KE KHD outer screw inner hole)
2 GND Negative power terminal
Note:
First, dB is A pure unit of count: for power, dB = 10*lg(A/B). For voltage or current, dB = 20*lg(A/B). The meaning of dB is actually very simple, is to represent a number that is very large (followed by a long string of zeros) or very small (preceded by a long string of zeros) relatively short. Such as:
X=1000000000000000 (15 zeros in total)
It can be expressed as: 10lgX=150dB
X = 0.000000000000001
It can be expressed as: 10lgX=-150 dB
dBm is defined as miliwatt(milliwatt). 0 dBm=1mw;
dBw defines watt. 0 dBW = 10lg1 W = 10lg1000 mw = 30 dBm.
dB is always defined as a unit of power by default, measured in 10lg.
Of course, in some cases, the signal strength (Amplitude) can be used to describe the work and power, at this time, 20lg is used as a measure. This is true in both control and signal processing.
For example, sometimes you can see the expression dBmV.
Package:
1pc HP353 RF Power Meter Programmable attenuation value Saving function 0.5G~15GHz -10~+16dBm