Specification
Chamber size(stainless steel SU304):
23 liter(Φ250mmX455mm)
Machine size:
23 liter(W:385mm H:365mm D:575mm)
Power: AC 110/1200W/11A;AC 220/1400W/6.3A
Sterilization temperature/sterilization pressure: 121℃-1.2bar(0.012Mpa) 134℃-2.2bar(0.022Mpa)
Sterilization time: 25Min(121℃) 6Min(134℃)
Lock system:
Double control with electric and mechanical lock
Guard against dry-combustion
Safety:
Over-heat protection Safety valve
Machine weight:
23 liter(G.W:33.8kg,N.W:32.3kg)
Sterilization Process:
“134℃ Program” button:
Heating to pressure 0.70bar(HE.1)→the first exhaust(TE.1)→Heating to pressure 1.70bar
(HE.2)→the second exhaust(TE.2)→Heating to pressure 2.20bar (HE.3)→ Start 135℃
Sterilization(Lasts 6 minutes)→the third exhaust(TE.3)→Balance to pressure0.00bar (LED display
“END”) → open the door (Sterilization completed).
“121℃ Program” button:
Heating to pressure 0.70 bar(HE.1)→the first exhaust(TE.1)→Heating to pressure 1.00bar (HE.2)
→the second exhaust(TE.2)→Heating to pressure 1.20bar (HE.3)→ Start 122℃ Sterilization(Lasts
25 minutes)→the third exhaust(TE.3)→Balance to pressure0.00bar(LED display “END”)→open
the door (Sterilization completed).
“Dry” Program button:
Caution: Dry program working needs the door is open.
Press “Dry” program button, the indicator starts to flash. Then press “START/STOP” button to start
19-minute dry program during indicator flashing.
“Exhaust” button :
When sterilizer was not running, but found high pressure inside. That’s need to use “Exhaust” button.
Operation: 1.Click “Exhaust” button. 2.Indicator will turn from “OFF” to “ON”. 3. Indicator turned
back to “ON”. 4. Door can be opened. 5.5 “Abnormally Exit”
“Abnormally Exit”:
Operation can be used during program running, but you want to stop.
Operation: 1. Hold on pressing “START/STOP” button 3 seconds, 2.Temperature window showing
flashing “E00”. And sterilizer started exhaust. 3.Click “START/STOP” again after pressure window
showing 0.0 bar with the sound KA from lock opened. 4.Door can be opened until temperature window showing
Package Listing:
Main Machine*1pc
Tray rack*1pc
tray*3pcs
Tray holder*1pc
Fuse*2pcs
Drain tube*1pc
Measuring cup*1pc
Warning:
1 Do not open the door of the sterilizer during operation and working!
2 Every time using needs minimal distilled water 1000ml.
3 In case of emergency, please disconnect the power immediately.
4 If there has leakage from the door, Please adjust the tightness to a proper level.