If not specified,
Default pairing with the largest size blanket 
(Sleeve belly window blanket,M200195,200*195)

Warming System utlizes the convection of air to transmit the hot air generated by the main unit to the disposable blanketthrough the pipeline, so thatthe heat energy is evenly distributed and transmited to the patient's body surface covered by the blanket to achieve the effect ofheat preservation olheating for prevention and treatment of hypothermia in patients. The blanket is made of non-woven fabric and mel-blown cloth delicate material. ltis with scientific air outlet arrangement, good temperature transter effect, it is widely used in Operation Room, lcU and other wards. Color LcD screerand quick setting button makes it user friendly.

Temperature Setting
32.0~41.0°C、95.0~105.8°F

Temperature Increment
0.1°C,0.1°F

Temperature Deviation
<± 1°C,<士1.8℉

Optional Unit
°C&°F

Warming Up Time
<4 minutes

Working Mode
Continuous working

Operation Time
Real time display

Alarm
Overheat, low temperature, no operation,fan error, temperature sensor error

Display
3.5-inch color LcD display

Air Volume
8 Level

Natural Wind Function
Available

Quick Temperature Setting
4Temp.

Alarm Volume
8 Level

Mute Time
2 minutes

History Record
10000pcs

Ambient Temperature Function
Available

Language
Chinese and English

Air Purification
Plasma air purification

Air Purification
Plasma air purification

Blanket
16 kinds disposable blankets optional

Filter
G4 (EN779) grade

Power Consumption
850VA

Power Supply
AC220V, 50Hz/60Hz

Type of Protection against Electric Shock
Class I

Degree of Protection against Electric Shock
Applied part Type BF protection against defibrillation discharge effects

Operation Environment
Temp:+10°C~+40°C;Humidity:30%~
75%; Atmospheric pressure: non-condensation,70KPa~106Kpa

Storage Environment
Temperature:-20°C~+55°C;Humidity:10~93%; Atmospheric pressure: non-condensation,50KPa~106Kpa

Degree of Protection against Ingress of Liquid
IPX2

Dimension 
260mmX325mmX325mm

Weight
≥6kg

Overheat Protection Function
Dual hardware and software protection, real-time temperature monitoring and over 50士5°C hardware forced power-off protection