CO2-M01 is used to match Patient Monitor. The module adopts sidestream method to measure the parameters related to ETCO2 and AWRR. It can detect pulmonary ventilation function and reflect the circulation and pulmonary blood flow and indirectly reflect the alveolar ventilation. 
The application population is adult patients. It is applicable for use in emergency department, ICU, operating room, respiratory department and other occasions. It is connected to the nose end of the patient through the sampling tube (applied part).

Features
1) End-tidal CO2 measurement function.
2) Inspired Minimum CO2 measurement function.
3) Respiration rate measurement function.
4) Optional three units: %, kPa, mmH.
5) Be compatible with Respironics module’s parameters and protocol.
6) No uesr calibration is required.
7) User can calibrate zero by themselves.
8) With atmospheric pressure and temperature compensation functions.

Performance
 End-tidal CO2 measurement:
1)Measurement range: 0mmHg ~150mmHg; 
2) Resolution: 1 mmHg;
3)Accuracy:
0~40 mmHg: ±2 mmHg
41~70 mmHg: ±5% of readings
71~100 mmHg: ±8% of readings
101~150 mmHg: ±10% of readings
 Respiration rate measurement:
1)Measurement range: 2rpm~120rpm;
2)Resolution: 1rpm;
3)Accuracy: ±1rpm;
 Inspired Minimum CO2 measurement:
Measurement range: 3mmHg ~50mmHg;
 Other parameters:
1)Preheat time: 2min;
2)Response time: < 3s;
3)Sampling frequency: 200 Hz.
 Safety characteristic
1)Degree of protection against electric shock: type BF applied part;
2)Degree of protection against ingress of liquid: IP22.

Accessory
1) A uers manual;
2) A CO2 Sampling tube.
3.5 Physical Characteristic
Dimension: 98.9*54.3*27.6 mm
Weight: 165 g
Working environment
Temperature: 5℃~40℃
Relative humidity: 30%~ 75%, no condensation
Atmospheric pressure: 700hPa~1060hPa
Power supply: DC 5.0V
Input power: ≤0.9VA

Optional configuration
1) Interface:5-pin (TTL);
          LEMO (RS232);
2) Accessory:
        Standard:
CO2 Sampling tube (Nasal cannula with PE water absorption filter element)
Optional:
CO2 Sampling tube (Nasal cannula with filter cotton and drying tube)
CO2 Sampling tube (Nasal cannula with filter cotton)
CO2 Sampling tube (Cannula tube with filter cotton and drying tube)
CO2 Sampling tube (Cannula tube with filter cotton)
CO2 Sampling tube(Cannula tube with PE water absorption filter element)

Buy safe Products
The following FDA Disclaimer is required for all eBay listing in Healthcare category and is included for REFERENCE:
The sale of this item may be subject to regulation by the U.S. Food and Drug Administration and state and local regulatory agencies.If the item is subject to FDA regulation, We will verify your status as an authorized purchaser of this item before shipping of the item.
If you have questions about legal obligations regarding sales of medical devices, you should consult with the FDA's Center for Devices and Radiological Health.
The Fingertip Pulse Oximeter is registered on the Australian Register of Therapeutic Goods (ARTG) with the code 197923, and certified by FDA of United States and CE,TUV of Europe.The  Fingertip Pulse Oximeter that is FDA 510K Approved

ibay365