Dentsply Sirona Restore Solution v2.1, a CAD/CAM software solution designed for Windows 7/10 PCs.
Dentsply Sirona Restore Solution v2.0
Objective
The objective of inserting the Restore Solution is to restore the software
factory defaults as a means of rectifying faults; for example:
● after a system crash,
● non-physical hard disk malfunctions,
● faulty software installations.
Compatible PCs
Dentsply Sirona Restore Solution v2.0 is compatible with the PCs of the
following devices:
● CEREC Primescan AC, Primescan AC
● CEREC Omnicam AC, Omnicam AC
● inLab 4 PC
● CEREC AC with CEREC Omnicam, CEREC AC Connect with
CEREC Omnicam
● CEREC AF / AI / AF Connect / AI Connect
2.2.3
Prerequisites for performing this procedure
● USB stick with Dentsply Sirona Restore Solution v2.0 or higher.
● All important user data on the C: drive must be backed up on the D: drive as a minimum; an even better and safer option is to back it up on an external medium, such as a CD, DVD, USB mass memory, or network.
● Up-to-date installation media of the software used on the systemmust be available (refer to the final steps).
● The hardware must be the same as at the time of delivery.
● The protected partition and the image file must not have been changed.
● The hard disk must be recognized by BIOS and have no physical malfunctions.
● There must be no changes made to the partitions on the hard disk.
When can the procedure not be used?
This procedure is not suited for the initial configuring of a new hard disk.
Do not start the process if the following occurs:
● System initialization or BIOS error messages are displayed.
● The monitor display is not present.
● Trackball or touchpad of the acquisition unit are not recognized.
● Drive C: contains important data for which no backup exists.
It is lost:
● All data, programs, updates, and settings which the user has stored, installed, and performed are no longer available following the procedure.
● Hardware components that are subsequently installed or modified
might not be recognized or may lead to error messages andmalfunctions.
Patient data usually saved on the D: drive is retained in this process.
For certainty, Dentsply Sirona also recommends saving this beforehand