This Agilent 1200 Quat Pump G1311A is a reliable and efficient tool for any laboratory or clinic. It has been thoroughly tested and is in perfect working order. This pump is ideal for use in chromatography systems, medical and lab equipment, devices, and healthcare applications. Additionally, it is suitable for use in solvent trays and has a Degasser 1322A attachment. The brand of this pump is Agilent, and the model number is G1311A. It is categorized under Chromatography Systems, Medical & Lab Equipment, Devices, Healthcare, Lab & Dental, and Business & Industrial.  Please see below for specifications:

Performance Specifications for Agilent 1200 Series Quaternary Pump 

Type                                                                    Specification 
Hydraulic system                                       Dual plunger in series pump with proprietary servo-controlled variable stroke drive,                                                                               floating plungers and active inlet valve 
Setable flow range                                      0.001 – 10 ml/min, in 0.001 ml/min increments 
Flow range                                                   0.2 – 10.0 ml/min 
Flow precision                                             £ 0.07% RSD, or £ 0.02 min SD whatever is greater, based on retention time at constant                                                                      room temperature 
Flow accuracy                                              ± 1% or 10 µl/min whatever is greater 
Pressure Operating range                           0 – 40 MPa (0 – 400 bar, 0 – 5880 psi) up to 5 ml/min 
Operating range                                           0 – 20 MPa (0 – 200 bar, 0 – 2950 psi) up to 10 ml/min Pressure pulsation < 2 %                                                                                   amplitude (typically < 1 %), at 1 ml/min isopropanol, at all pressures > 1 MPa (10 bar) 
Compressibility compensation                    User-selectable, based on mobile phase compressibility 
Recommended pH range                            1.0 – 12.5, solvents with pH < 2.3 should not contain acids which attack stainless steel 
Gradient formation                                       Low pressure quaternary mixing/gradient capability using proprietary high-speed                                                                                   proportioning valve 
Delay volume                                               800 – 1100 µl, dependent on back pressure Composition Range 0 – 95 % or 5 – 100 %,                                                                       user selectable 
Composition Precision                                < 0.2 % RSD, at 0.2 and 1 ml/min 
Control and data evaluation                         Agilent control software (e.g. ChemStation, EZChrom, OL, etc.) 
Communications                                          Controller-area network (CAN), GPIB, RS-232C, APG Remote: ready, start, stop and                                                                           shut-down signals, LAN optional  
Safety and maintenance                              Extensive diagnostics, error detection and display (through control module and Agilent                                                                         Lab Monitor & Diagnostic Software), leak detection, safe leak handling, leak output                                                                               signal for shutdown of pumping system. Low voltages in major maintenance areas. 
GLP features                                                Early maintenance feedback (EMF) for continuous tracking of instrument usage in terms                                                                       of seal wear and volume of pumped mobile phase with user-settable limits and feedback                                                       messages. Electronic records of maintenance and errors. 
Housing                                                        All materials recyclable